Will USB-6008 DAQ be able to power and record voltage for UMS T5 tensiometer at the same time?

$
0
0

I would like to use my NI USB-6008 to provide power to my UMS T5 tensiometer  (http://www.ums-muc.de/en/products/tensiometer/t5.html) as well as take readings from it, but I don't know if this is possible to do properly. The supply voltage for the instrument can be as low as 5V, which I can easily get from the dedicated +5V channel. I am able to power the instrument and connect it to an analog input on the 6008 and measure a voltage in differential mode. However, while reading the support literature for the instrument, I find the following:

"Potential data acquisition pitfalls: Because the pressure transducer is configured in a Wheatstone full bridge, the input voltage and signal mV output cannot be connected to the same reference (ground). Hence, the signal mV output can only be measured using a differential voltage measurement. Therefore, do not attempt to make a single-ended measurement of the pressure transducer mV output."  (http://www.decagon.com/assets/Uploads/MeasuringUMSTensiometerswithnon-UMSControlandDataAcquisitionSystems.pdf)

My understanding is that the 6008 can take a differential measurement if I attach the signal "+" and signal "-" to the positive and negative analog input terminals. However, it seems that all of the Ground ports on the 6008 are grounded to the same reference, which would make my voltage measurement invalid according to the paragraph above. So my real questions is: if I attempt to record the voltage using one of the analog inputs on the 6008 in this way, is the measurement valid? Or do I need to find a separate power supply with a different reference ground to ensure that the measurement is accurate?


Viewing all articles
Browse latest Browse all 6878

Trending Articles