Hello,
I'm running a cDAQ-9174 which is populated with 4 cards (2x 9207 in slots 1 & 2, 9265 slot 3, 9485 slot 4). I was attempting to configure the system to measure analog inputs across the two 9207 cards using MAX. With nothing connected, I found that all of the Analog Ins on card 1 were reading close to 0V while all of the Analog Ins on card 2 were clipped at 10.4V. I swapped the cards between slots 1 & 2 and the 10.4V signal follows the board from slot 2 in the slot 1 position. Do I have a bad board? The system has never been connected to an input and passes all of the "self tests". The trend is the same regardless of if I try to make a voltage task or use the test panel. Interestingly, even if I switch the max input range for the card in the test panel to 0-5V it still clips at 10.4V.
Thanks,
Justin